- BEMC
- BEMC Detector Operator Manual
- Calibrations
- BPRS
- Run 8 BPRS Calibration
- 01 DB peds R9069005, 9067013
- 02 pedestal(capID)
- 03 tagging desynchronized capID
- 04 BPRS sees beam background?
- 05 ---- peds(softID,capID) & status table, ver=2.2, R9067013
- 06 MIP algo ver 1.1
- 07 BPRS peds vs. time
- 08 BPRS ped calculation using average
- 09 BPRS swaps, IGNORING Rory's finding from 2007, take 1
- 10 -------- BPRS swaps take2, _AFTER_ applying Rory's swaps
- 11 BPRS absolute gains from MIP, ver1.0 ( example of towers)
- 12 MIP gains ver1.0 (all tiles, also BTOW)
- 13 MIP algo, ver=1.1 (example of towers)
- 14 ---- MIP gains ver=1.6 , 90% of 4800 tiles ----
- 15 Broken BPRS channels ver=1.6, based on data from March of 2008
- 16 correlation of MIP ADC vs. raw slopes
- Run 9 BPRS Calibration
- Run 8 BPRS Calibration
- BSMD
- BTOW - Calibration Procedure
- Run 12 BTOW Calibration
- Run 3 BTOW Calibration
- Run 4 BTOW Calibration
- Run 5 BTOW Calibration
- Run 6 BTOW Calibration
- Run 7 BTOW Calibration
- Run 8 BTOW Calibration (2008)
- Run 9 BTOW Calibration
- BPRS
- Database
- Hardware
- Mapping
- Service Tasks
- Software
- Useful Documents
- EEMC
- ETOF
- FCS
- FGT
- FPD & FMS & FPS
- FTPC
- FTT
- HLT
- L3
- PMD
- PP2PP
- RICH
- Roman Pot Phase II*
- SSD
- SVT
- Slow Controls
- TPC
- TRG
- Trigger Detectors
- VPD
- test
01 DB peds R9069005, 9067013
Updated on Tue, 2009-03-10 14:33. Originally created by mattheww on 2008-10-15 17:20.
Under:
Pedestal residua for 434 zero-bias events from run 9069005.
The same pedestal for all caps was used - as implemented in the offline DB.
Fig 1.
Fig 2. Run 9067013, excluded caps >120. All 4800 tiles, pedestal residua from 100 st_zeroBias events. Y-axis [-50,+150].
Fig 3. Run 9067013, excluded caps >120. Pedestal corrected spectra for all 4800 tiles, 10K st_physics events. Y-axis [-50,+150].
Dead MAPMT results with 4 patches 4 towers wide.
Fig 4.
Run 9067013, excluded caps >120.
Zoom-in Pedestal corrected spectra, one ped per channel.
Top 10K st_physics events (barrel often triggered)
Bottom pedestal residua 100 st_zeroBias events
Fig 5.
Run 9067013, input =100 events, accept if capID=124 , raw spectra.
There are 4 BPRS crates, so 1200 channels/crate. In terms of softIds it's
PSD1W: 1-300 + 1501-2400
PSD19W: 301-1500
PSD1E: 2401-2900 + 4101-4800
PSD20E: 2901-4100
Why only 2 channels fired in crate PSD20E ?
»
- Printer-friendly version
- Login or register to post comments